DCT 400A TRANSDUCERS

Current Sensing - Flux-Gate Transducers - DCT 400A Transducers

Features:

  • High Accuracy over high bandwidth.
  • Very low output noise and offset drift.          
  • Negligible insertion losses.          
  • High immunity to interference.          
  • Overload Capability.          
  • Excellent linearity.  

 

General Data:  

  • Operating Temperature: -20 to 70 ºC 
  • Storage Temperature: -20 to +85 ºC 
  • Weight: 400g
  • Primary Diameter Hole: 25mm
  • Basic Insulation (Between Primary and Measurement Current): 3500V AC 50Hz 1' 

 

 Accuracy:  

  • Accuracy at Ip (T=25ºC): < 0.1%
  • Linear Error ( Beetwen 100 to 400 A) Vcc = ±15V, Rb = 1.8 Ohm: < 5 ppm 
  • Linear Error ( Beetwen 10A to 100A ) Vcc = ±15V, Rb = 10 Ohm: < 10 ppm 
  • Linear Error ( Beetwen 0.1 to 10A ) Vcc = ±15V, Rb = 33 Ohm: < 50 ppm 
  • Offset Current: 5uA Max
  • Offser Current Temperature Drift: < 5 ppm/ºC 
  • Time Response ( 10% to 90% of Ip): <1us 
  • di/dt Followed Accurately: > 100A/us
  • Frequency Bandwidth ( Ip = 10A DC): DC to 100kHz ( -3dB ) 

 

 Output Inputs Connector Information: 

  • Connector: DB-9 Standard Type (Female)
  • Fault Operation Condition (Led Power OFF): Ip > 120 %
  • Maximum Switching Current (pins 3 to 8 and pin 3 to 7): 2A
  • Maximum Switching Voltage (pins 3 to 8 and pin 3 to 7): 30 VDC/120 VAC
  • Compensation Winding Maximum Resistance (T = 50ºC): 30 Ohm RC
  • Length Two Wire Cable to R burden (Connected beetwen pin 1 to 6): 50cm (Typical)

  

Applications:     

  • Precise and high stability inverters.          
  • Energy measurements.          
  • High Precision Power Supplies.          
  • Feed back element in high performance gradient amplifiers for MRI.  
  • Medical Equipment.  

 

 

According to:

  • UNE EN 50178
  • UNE EN 50155